Entwicklung von Atmosphärendruck- Ionenquellen für die Massenspektrometrie
High throughput LC-MS is an increasing topic in analytical chemistry. Especially the idle time of a mass spectrometer should be reduced to use these devices more efficiently and save costs. Therefore, the first section of this thesis demonstrated how a dual ion source could be constructed. This ion source was designed to operate the typical ionization modes of LC-MS analyses, ESI, and APCI simultaneously and individually. Direct infusion experiments of reserpine and caffeine indicated that it is possible to run the two ion sources simultaneously. Furthermore, it was shown that switching between the two ion sources does not lead to instabilities in the ion current of the mass spectrometer. A six-port valve was used to couple a HPLC to the dual ion source for the initial chromatographic studies. Switching experiments were performed with cortisol and ß-estradiol, resulting in a minimum switching time between the two probes of 6 s. Repeatability was investigated to be 0.28% intraday and 1.5% interday for APCI and 2.2% intraday and 2.8% interday for ESI. The limit of detection for the standard substances testosterone (ESI), vitamin D3 (APCI), and 25-hydroxyvitamin D3 (ESI) was 30 ng L-1 for testosterone and vitamin D3 and one µg L-1 for 25-hydroxyvitamin D3. The dual ion source was coupled to two HPLCs, making it possible to shorten the idle time of the mass spectrometer and to use individual methods for the two HPLCs. With this technique, the LOD of vitamin D3 could be reduced to 300 ng L-1. The final experiment for multiple switching between ESI and APCI and the two UHPLCs shows good repeatability of the peak areas with standard deviations of 1.5% for testosterone and 3.8% for vitamin D3. In five runs, 50 analyses could be performed, requiring 1.5 min run time for one sample. Therefore, the dual ion source is well suited for the coupling of multiple HPLCs, accompanied by a low analysis time and a low idle time of the mass spectrometer.