Diagnostic accuracy of 18F–FDG PET/CT and MR imaging in patients with adenoid cystic carcinoma

Ruhlmann, Verena LSF; Pöppel, Thorsten D. LSF; Veit, Johannes; Nagarajah, James LSF; Umutlu, Lale LSF; Hoffmann, Thomas K.; Bockisch, Andreas LSF; Herrmann, Ken LSF; Sauerwein, Wolfgang LSF

Background:
The aim of this study was to evaluate the value of 18F–FDG PET/CT (PET/CT) and MRI for local and/or whole-body restaging of adenoid cystic carcinoma of the head and neck (ACC).
Methods:
Thirty-six patients with ACC underwent conventional MRI of the head and neck and a whole-body PET/CT and were analysed with regards to detection of a local tumor recurrence, lymph node or distant metastases. A consensus interpretation of all available imaging data was used as reference standard. Sensitivity, specificity, diagnostic accuracy, positive and negative predictive values were calculated for MRI and PET/CT.
Results:
The sensitivity of PET/CT and MRI was 96% (89%), specificity 89% (89%), PPV 96% (96%), NPV 89% (73%) and accuracy 94% (89%) for detection of local tumors. Additionally, PET/CT revealed lymph node metastases in one patient and distant metastases in 9/36 patients. In three patients secondary primaries were found.
Conclusions:
Whole-body PET/CT in addition to MRI of the head and neck improves detection of local tumour and metastastic spread in ACC.